Sample
|
Whole blood, Serum, or Plasma
|
Sample Size
|
100
ml Whole Blood, Serum or Plasma
400
ml
diluted (1:10) Urine
|
Detection
Range
|
Blood:
Na+: 20-200 mmol/L
K+: 0.2-40 mmol/L
Li+: 0.0-5.0 mmol/L
Urine:
Na+: 25-1000
mmol/L (undiluted)
K+: 1.0-500 mmol/L (undiluted)
|
Analysis Time
|
60 sec (Blood)
90 sec (Urine)
|
Calibration
|
Automatic or On-Demand
|
Resolution
|
Na+:
0.10 mmol/L
K+: 0.01 mmol/L
Li+: 0.01 mmol/L
|
Reproducibility
|
Typical Within Run (n=20)
Blood, Serum, Plasma
Na+: CV < 1% (80-200 mmol/L)
K+:
CV < 2% (1-10 mmol/L)
Li+: SD < 0.03 mmol/L (0.2-2.0 mmol/L)
Urine
Na+: CV < 5% (80-200 mmol/L)
K+:
CV < 5% (1-10 mmol/L)
|
Data Storage
|
100 patient results
20 normal QC results
20 abnormal QC results
|
Accuracy
|
Typical vs Flame Photometer
Blood, Serum, Plasma
Na+: r= 0.99 (80-200 mmol/L)
K+:
r= 1.00 (1-10 mmol/L)
Li+: r= 0.99(0.2-2.0 mmol/L)
|
Output
|
16 place alphanumeric display
40 column printer
Serial port (RS-232)
AutoSampler port
|
Ambient Conditions
|
15-32C (60-90F), <85% humidity
|
Power
|
100-115 ~VAC 50/60 Hz, 0.8A
or 220 ~VAC 50/60 HZ, 0.4A (Factory set)
|
Size and Weight
|
9.5 in. (24 cm) W x 16.5 in. (42cm)
H x 8.0 in. (20 cm) D, 13 lbs. (5.8 kg)
|