EasyLyte Na/K/Li Specifications
Sample
Whole blood, Serum, or Plasma 
   

Sample Size

100 ml Whole Blood, Serum or Plasma
4
00 ml diluted (1:10) Urine
  

Detection Range 

Blood:
Na+: 20-200 mmol/L
K+: 0.2-40 mmol/L
Li+: 0.0-5.0 mmol/L
Urine:
Na+: 25-1000 mmol/L (undiluted)
K+: 1.0-500 mmol/L (undiluted)

  

Analysis Time

60 sec (Blood)
90 sec (Urine)

  

Calibration

Automatic or On-Demand
  

Resolution

Na+: 0.10 mmol/L
K+: 0.01 mmol/L
Li+: 0.01 mmol/L

  

Reproducibility

Typical Within Run (n=20)
Blood, Serum, Plasma
Na+: CV < 1% (80-200 mmol/L)
K+: CV < 2% (1-10 mmol/L)
Li+: SD < 0.03 mmol/L (0.2-2.0 mmol/L)
Urine
Na+: CV < 5% (80-200 mmol/L)
K+: CV < 5% (1-10 mmol/L)

  

Data Storage

100 patient results
20 normal QC results
20 abnormal QC results
  

Accuracy

Typical vs Flame Photometer
Blood, Serum, Plasma
Na+: r= 0.99 (80-200 mmol/L)
K+: r= 1.00 (1-10 mmol/L)
Li+: r= 0.99(0.2-2.0 mmol/L)

   

Output

16 place alphanumeric display 
40 column printer
Serial port (RS-232)
AutoSampler port
  

Ambient Conditions

15-32C (60-90F), <85% humidity
  

Power

100-115 ~VAC 50/60 Hz, 0.8A 
or 220 ~VAC 50/60 HZ, 0.4A (Factory set)

  

Size and Weight

9.5 in. (24 cm) W x 16.5 in. (42cm) 
H x 8.0 in. (20 cm) D, 13 lbs. (5.8 kg)
  