EasyLyte Na/K/Cl/Li Specifications
Sample
Whole Blood, Serum, Plasma, or Urine 
   

Sample Size

100 ml Whole Blood, Serum or Plasma
400 ml diluted (1:10) Urine
    

Detection Range 

Blood:
Na+: 20-200 mmol/L
K+: 0.2-40 mmol/L
Cl-: 25-200 mmol/L
Li+: 0.2-5.0 mmol/L

Urine:
Na+: 25-1000 mmol/L (undiluted)
K+: 1.0-500 mmol/L (undiluted)
Cl-: 25-500 mmol/L (undiluted)
Li+: (N/A)

  

Analysis Time

55 sec (Blood)90 sec (Urine)
  

Calibration

Automatic or On-Demand
  

Resolution

Na+: 0.10 mmol/L
K+: 0.01 mmol/L
Li+: 0.01 mmol/L

  

Reproducibility

Typical Within Run (n=20)
Blood, Serum, Plasma:
Na+: CV < 1% (80-200 mmol/L)
K+: CV < 2% (1-10 mmol/L)
Cl-: CV < 2% (80-200 mmol/L)
Li+: SD < 0.03 mmol/L (0.2-2.0 mmol/L)
Urine:
Na+: CV < 5% (80-200 mmol/L)
K+: CV < 5% (1-10 mmol/L)
Cl-: CV < 5% (80-200 mmol/L)
Li+: (N/A)

  

Data Storage

125 patient results
Three levels of QC results
  

Output

32 place alphanumeric display
40 column printer
Serial port (RS-232)
  

Ambient Conditions

15-32C (60-90F), <85% humidity
  

Power

100-115 ~VAC 50/60 Hz, 0.8 A 
or 220 ~VAC 50/60 HZ, 0.4A
  

Size and Weight

9.5 in. (24 cm) W x 16.5 in. (42cm) H x 8.0 in. (20 
cm) D, 13 lbs. (5.8 kg)
  