Parameters
|
Measures Na+, K+, Ca++, pH
Measures Ca++, pH at 37C + 0.1C
Ionized Calcium reported at sample pH
|
Sample
|
Whole Blood, Serum or Plasma
|
Sample
Size
|
100
ml Whole Blood, Serum or Plasma
|
Detection
Range
|
Blood:
Na+: 20-200 mmol/L
K+: 0.2-40 mmol/L
Ca++: 0.1-6 mmol/L
pH:
6.0-8.0 pH units
|
Analysis Time
|
60 sec (Blood)
|
Temperature of Measurement
|
37C+0.1C |
Analytes
Calculated and Reported
|
Ca++ at pH = 7.4
|
Calibration
|
Automatic or On-Demand
|
Resolution
|
Na+: 0.10 mmol/L
K+: 0.01 mmol/L
Ca++: 0.01 mmol/L
pH: 0.005 pH units
|
Reproducibility
|
Typical Within Run (n=20)
Blood, Serum, Plasma
Na+: CV< 1% (80-200 mmol/L)
K+: CV< 2% (1.0-10.0 mmol/L)
Ca++: SD<0.02 (0.4-1.5 mmol/L)
pH: SD<0.01 (6.8-8.0 pH units)
|
Data Storage
|
125 patient results
Three levels of QC results
|
Output
|
32 character alphanumeric display
40 column printer
Serial port (RS-232)
|
Ambient Conditions
|
15-32C (60-90F), <85% humidity
|
Power
|
100-115 ~VAC 50/60 Hz, 0.8 A
or 220 ~VAC 50/60 Hz, 0.4 A
|
Size and Weight
|
9.5 in. (24 cm) W x 16.5 in. (42cm) H x 8.0 in. (20 cm)