EasyLyte Na/K Specifications

Sample
Whole Blood, Serum, Plasma, or Urine
   

Sample Size

100 ml Whole Blood, Serum or Plasma
60 ml Capillary Sample
400
ml
diluted (1:10) Urine
   

Detection Range 

Blood:
Na
+: 20-200 mmol/L
K
+: 0.2-40 mmol/L
Urine:
Na
+: 25-1000 mmol/L (undiluted)
K
+: 1.0-500 mmol/L (undiluted)
  

Analysis Time

60 sec (Blood)90 sec (Urine)
  

Data Storage

100 patient results
20 normal QC results
20 abnormal QC results
  

Calibration

Automatic or On-Demand
  

Resolution

Na+: 0.10 mmol/L
K
+: 0.01 mmol/L
  

Reproducibility

Typical Within Run (n=20)
Blood, Serum, Plasma:
Na
+: CV < 1% (80-200 mmol/L)
K
+: CV < 2% (1-10 mmol/L)
Urine:
Na
+: CV < 5% (80-200 mmol/L)
K
+: CV < 5% (1-10 mmol/L)
  

Output

16 place alphanumeric display
40 column printer
Serial port (RS-232)
AutoSampler port
  

Ambient Conditions

15-32 C (60-90 F), <85% humidity
  

Power

100-115 ~VAC 50/60 Hz, 0.8 A 
or 220 ~VAC 50/60 HZ, 0.4 A (Factory set)
  

Size and Weight

9.5 in. (24 cm) W x 16.5 in. (42  cm) H x 8.0 in. (20 cm) D, 13 lbs. (5.8 kg)
  